Product Details:
|
| Type: | Testing Machine | Accuracy Class: | High Accuracy |
|---|---|---|---|
| Accuracy: | / | Application: | Auto Testing |
| Customized Support: | OEM, ODM, OBM | Power: | --- |
| Protection Class: | Ip56 | Voltage: | 220V Electrical Equipment |
| Warranty: | 1 Year | Current Consumption: | 30mA |
| Power Supply: | AC220V±10% 50Hz±5% | Instrument Weight: | 5kg |
| Dimensions: | 224*110*40mm | ||
| Highlight: | ISO 5165 octane number tester,ASTM D2699 cetane analyzer,gasoline octane test machine |
||
ASTM D2699 EN ISO 5165 Octane Number Tester Cetane Number Analyzer ASTM D2700 ASTM D4737 Gasoline Octane Number Test Machine
Product Introduction
The octane number and cetane number analyzer adopts international technical measurement principles. It is designed and developed based on the diversity of domestic oil products, filling the gap in the detection of various oil types. It can quickly detect the octane number of gasoline and the cetane number of diesel. It has digital displays in both Chinese and English. It has a large capacity for data storage and can directly print the measured values. The page design is user-friendly and simple to operate. It uses original imported shells and chips. The sensors are designed with a full-sealed alloy metal structure. It has an ultra-long standby cycle charging function. It is suitable for work sites, outdoors, laboratory operations. It is widely used in the fields of petroleum, chemical industry, power, commodity inspection, universities, research institutions, etc.
Functional Features
1. The gasoline octane number and diesel cetane number testing instrument can be used for both purposes in one device. It is portable and easy to carry.
2. After long-term data collection, the database is complete. By integrating new technologies, multiple detection modes have been developed. It can detect various types of vehicle gasoline, ethanol gasoline, high-octane oil products, low-octane oil products, methanol gasoline (only providing reference values), methanol diesel, catalytic diesel, biodiesel, emulsified diesel, and other types of oil products. This fills the gap in the detection of octane value and cetane value for these oil products.
3. Introducing the international dielectric constant measurement principle, it has been designed and developed based on the demands of the Chinese market. It is an instrument that fully meets the requirements of the domestic market and incorporates the advantages of the original domestic testing instruments.
4. The detection is fast and accurate, with simple operation. It features a large screen with Chinese display, and is equipped with an EL backlight for enhanced visibility. It is convenient to use in dimly lit environments. The main panel offers a Chinese menu-style operation.
5. Original imported instrument casing, sturdy and durable, and compact and easy to carry. The chips and core components are all imported, which enhances the instrument's sensitivity and service life. The sensor alloy metal is fully sealed. The outer packaging box has high strength, high impact resistance, waterproof, moisture-proof, dust-proof, and is equipped with a grid sponge inside.
6. The upper and lower limits of the mode can be preset in advance. When exceeding the range, it will automatically display, making it convenient for users to check the required values.
7. Mode curves can be set to calibrate the instrument. You can choose either 2-point or 3-point calibration.
8. It has an automatic power-off energy-saving function. With a low-power design, it uses 4 rechargeable batteries that can be charged repeatedly. The batteries are more reliable.
9. Chinese/English interface display (can be selected freely)
ASTM D2699 EN ISO 5165 Octane Number Tester Cetane Number Analyzer ASTM D2700 ASTM D4737 Gasoline Octane Number Test Machine
Technical Specification
|
Adaptation Standard |
ASTM D2699-86, ASTM D2700-86, ASTM D 4737-03, ASTM D 613, EN ISO 5165 |
|
Octane Number Measurement Range |
40-120(ON) |
|
Cetane Number Measurement Range |
20-100 (CN) |
|
Operating Method |
Key operation, LCD display |
|
Octane Number Error |
±0.5 |
|
Cetane Number Error |
±1 |
|
Octane Number Repeatability |
±0.2 |
|
Cetane Number Repeatability |
±0.5 |
|
Calibration Point |
2 points, 3 points |
|
Measurement Time |
≤30s |
|
Battery Overvoltage Threshold |
2.5 volts |
|
Usage Time |
15h |
|
Display |
Liquid crystal display |
|
Print Function |
Automatic printing |
|
Current Consumption |
30mA |
|
Environmental Temperature |
-15℃ to 45℃ |
|
Relative Humidity |
10% to 80% Rh |
|
Sensor Size |
60×110mm |
|
Dimensions |
224×110×40mm |
|
Instrument Weight |
5kg |
![]()
![]()
![]()
![]()