Product Details:
|
| Type: | Testing Machine | Accuracy Class: | High Accuracy |
|---|---|---|---|
| Accuracy: | ±0.2% | Application: | Auto Testing, Laboratory Testing, Lab Testing |
| Customized Support: | OEM, ODM, OBM, Software Reengineering | Power: | --- |
| Protection Class: | Ip56 | Voltage: | 220 V |
| Warranty: | 1 Year | Product Name: | Specific Heat Capacity |
| Highlight: | low temperature specific heat capacity tester,specific heat capacity analyzer,thermal analyzer test method |
||
Low temperature specific heat capacity tester/Test method for specific heat capacity/Specific heat capacity analyzer
![]()
The instrument refers to GJB 330A-2000 solid material 60-273K specific heat capacity test method (60-373K adiabatic method) design and manufacture. All measurement and control are controlled by computer to reduce human error and improve testing accuracy. The automatic control, data acquisition and processing, display and print-out of the instrument are realized by computer interface operation.
The instrument adopts adiabatic method to test the specific heat capacity of solid (powdery, small piece or small piece) metal and non-metal materials at low temperature, widely used in production enterprises, quality inspection departments, universities and research institutes and other scientific research units.
Two, Xiangke BRR-DW low temperature specific heat capacity tester main technical indicators:
1, measuring temperature range: -40 -- 100℃, 0 -- 100℃ and other specifications optional.
2, specific heat capacity range: 0.05-5(KJ /kg* K); Accuracy of 1% plus or minus 0.008 or less
3, sample size: φ30×80 (mm).
4, calorimeter temperature measurement resolution: 0.001℃.
5, calorimeter power supply: voltage 0-30V, resolution 0.1mV; Current 0-1A, resolution 0.01mA.
6, the instrument with vacuum device, vacuum degree: 0.09mpa.
7, insulation screen: internal insulation screen and external insulation screen are using 4 pairs of differential thermopile, temperature resolution 0.01/4℃.
8, temperature control mode: constant temperature tank
9, computer automatic test control.
10, the instrument power supply: 220V/50Hz, power consumption is less than 1KW.
Instrument configuration:
1. One BRR-DW test host,
2. One copy of test software
3, a low temperature constant temperature tank,
4. One vacuum pump,
5. One desktop computer,
6. One set of communication interface and data line.
![]()
![]()
![]()
![]()
Contact Person: Kaitlyn Wang
Tel: 19376687282
Fax: 86-769-83078748